smithery/gptomics

bio-flow-cytometry-bead-normalization

Bead-based signal normalization and cross-batch harmonization for CyTOF and high-parameter cytometry - EQ four-element bead normalization of instrument sensitivity drift (CATALYST normCytof, premessa), and reference-anchor cross-batch normalization (CytoNorm, per-cluster quantile splines). Covers the distinction between within-run drift correction and between-batch correction, the mandatory anchor/reference sample, why normalization is per-cluster with many quantiles, and the over-correction ri…

Installation

$ npx skills add smithery/gptomics --skill bio-flow-cytometry-bead-normalization

Summary

  • Bead-based signal normalization and cross-batch harmonization for CyTOF and high-parameter cytometry - EQ four-element bead normalization of instrument sensitivity drift (CATALYST normCytof, premessa), and reference-anchor cross-batch normalization (CytoNorm, per-cluster quantile splines).
  • Covers the distinction between within-run drift correction and between-batch correction, the mandatory anchor/reference sample, why normalization is per-cluster with many quantiles, and the over-correction risk.
  • Use when correcting CyTOF signal drift, harmonizing multi-batch or multi-site studies, or deciding whether to normalize data versus model batch in the design.

Similar popular skills

Related neighbors and high-traction skills in the same topics — useful to compare before installing.

Also in this package

Other skills from smithery/gptomics · top by installs.

npx skills add smithery/gptomics

Browse all from smithery/gptomics

More details

Agent compatibility

Declared targets from SKILL.md / docs. Unmarked agents are not listed — the skill may still install via the CLI.

Claude Code Not declared
Cursor Not declared
Codex Not declared
GitHub Copilot Not declared
Windsurf Not declared
Gemini CLI Not declared
Cline Not declared
OpenCode Not declared

Package contents

Files included with this skill beyond the listing page.

  • skill md SKILL.md 8,026 B
  • docs SUMMARY.md 276 B

History

  1. First recorded snapshot · 0 installs

SKILL.md

Version Compatibility

Reference examples tested with: CATALYST 1.26+, CytoNorm 2.0+, flowCore 2.14+.

Before using code patterns, verify installed versions match. If versions differ:

  • R: packageVersion('<pkg>') then ?function_name to verify parameters

normCytof() returns a LIST ($data, $beads, $removed, ...), not a flowFrame; beads="dvs" encodes EQ masses 140,151,153,165,175. Confirm with ?normCytof before relying on slot names.

Bead Normalization

"Normalize my CyTOF data" -> Correct instrument sensitivity drift with EQ beads (within/across runs), then harmonize batches with a reference anchor.

  • R (drift): CATALYST::normCytof() (EQ-bead-based) or premessa
  • R (batch): CytoNorm::CytoNorm.train() + CytoNorm.normalize() (per-cluster quantile splines)

The Single Most Important Modern Insight -- Two Different Layers; Anchor Controls Are the Guarantee

Bead normalization and batch normalization correct DIFFERENT things and are NOT interchangeable. (1) EQ-BEAD normalization (Finck 2013 Cytometry A 83:483) corrects within-run and run-to-run instrument SENSITIVITY DRIFT using the four-element beads as a physical internal standard - applied first, on raw counts. (2) CROSS-BATCH normalization (CytoNorm, Van Gassen 2020 Cytometry A 97:268) corrects staining/acquisition batch effects using a shared ANCHOR/reference sample present in EVERY batch, learning per-FlowSOM-cluster quantile-spline transforms. Beads cannot fix staining-batch or reagent-lot effects; CytoNorm cannot fix intra-run detector drift. The anchor control is the load-bearing design element: because it is biologically identical across batches, any cross-batch difference in it is technical BY CONSTRUCTION. Dropping the anchor (CytoNorm 2.0) is convenient but reintroduces the over-correction risk the anchor was designed to eliminate - so the safest stance for inference is to MODEL batch in the diffcyt design and reserve normalization for visualization/clustering display.

Why Per-Cluster and Many (99) Quantiles

Batch effects are cell-type-specific - a marker can drift in monocytes but not in T cells - so a single global channel transform over-corrects one population while under-correcting another and can erase real abundance differences. CytoNorm therefore learns the transform PER FlowSOM cluster. And it uses ~99 quantiles + a spline because the drift is non-linear and intensity-dependent (the negative and positive peaks move by different amounts); a single median shift or linear rescale reintroduces the distortion it is trying to remove.

EQ-Bead Normalization (drift)

Goal: Correct sensitivity drift and remove bead events.

Approach: normCytof() gates beads, computes the correction on the linear scale, and returns a list - the cleaned SCE is in $data.

library(CATALYST)

sce <- prepData(fs, panel, md)                         # no by_time arg - normalization is normCytof's job
res <- normCytof(sce, beads = 'dvs',                   # EQ masses 140,151,153,165,175
                 k = 500, remove_beads = TRUE, overwrite = FALSE)   # k = smoothing window (default; affects bead-trace viz, not correction magnitude)
sce_norm <- res$data                                   # normalized SCE; res$beads / res$removed available

Cross-Batch Normalization (CytoNorm)

Goal: Harmonize batches using a shared reference sample.

Approach: Train on the anchor (present in every batch) -> learn per-cluster quantile splines -> apply to the real samples. testCV() first: if cluster CV is high, the FlowSOM model is batch-unstable and per-cluster splines will distort (fall back to nClus=1).

library(CytoNorm)

model <- CytoNorm.train(files = ref_files, labels = batch_labels, channels = marker_channels,
                        transformList = tl,
                        FlowSOM.params = list(nCells = 6000, xdim = 10, ydim = 10, nClus = 10),
                        normMethod.train = QuantileNorm.train,
                        normParams = list(nQ = 99), seed = 42)
CytoNorm.normalize(model = model, files = sample_files, labels = batch_labels,
                   transformList = tl, transformList.reverse = tl_rev,   # BOTH required
                   outputDir = 'normalized/')

Per-Method Failure Modes

Treating bead and batch normalization as the same

Trigger: expecting beads to fix staining-batch effects. Mechanism: different layers. Symptom: residual batch structure after bead norm. Fix: bead norm for drift; CytoNorm for batch.

No anchor in a batch

Trigger: a batch lacking the reference sample. Mechanism: nothing biologically-identical to learn from. Symptom: that batch can't be normalized / is over-corrected. Fix: run the anchor in every batch (or model batch instead).

Over-correction

Trigger: CytoNorm with groups confounded with batch, or anchor-free on variable samples. Mechanism: splines absorb real biology. Symptom: attenuated group differences. Fix: testCV() check; model batch in diffcyt for inference; normalize for display only.

Using normCytof return as a flowFrame

Trigger: sce_norm <- normCytof(...). Mechanism: it returns a list. Symptom: downstream type error. Fix: res$data.

Quantitative Thresholds

Threshold Source Rationale
bead drift reduced ~4.9x -> 1.3x Finck 2013 Cytometry A 83:483 EQ-bead correction over a month of runs
99 quantiles, per-cluster Van Gassen 2020 Cytometry A 97:268 non-linear intensity-dependent, cell-type-specific drift
EQ masses 140,151,153,165,175 (dvs) CATALYST DVS/Fluidigm EQ four-element bead set

Common Errors

Error / symptom Cause Solution
normCytof output not usable it returns a list use res$data
prepData(by_time=TRUE) errors no such argument use normCytof() for bead/drift correction
CytoNorm distorts populations unstable FlowSOM clustering run testCV(); reduce nClus (or 1)
batch effect remains only bead-normalized add CytoNorm with anchor samples

References

  • Finck 2013 Cytometry A 83(5):483-494 — EQ-bead normalization of CyTOF drift.
  • Van Gassen 2020 Cytometry A 97(3):268-278 — CytoNorm per-cluster quantile normalization.
  • Quintelier 2025 Cytometry A 107(2):69-87 — CytoNorm 2.0 (anchor-free; over-correction caveat).
  • Chevrier 2018 Cell Syst 6(5):612-620 — CyTOF spillover (CATALYST normalization context).

Related Skills

Workflow order (CyTOF): EQ-bead drift normalization (raw counts, FIRST) -> cytometry-qc -> doublet-detection -> clustering -> CytoNorm cross-batch (LAST). The two normalization layers sit at opposite ends.

  • cytometry-qc - EQ-bead-median-vs-Time is the primary CyTOF drift readout
  • doublet-detection - Remove doublets before normalization
  • compensation-transformation - Transform scale used by CytoNorm
  • clustering-phenotyping - Cluster across normalized batches
  • differential-analysis - Model batch in the design rather than over-cleaning
  • experimental-design/batch-design - Anchor/reference-sample design; differential-expression/batch-correction for execution